PERFORMING A CONTROL SOLUTION TEST

Complete steps 1-6, first using Level 1 control solution, again using Level 2 control solution.

1

2

Step 1: To turn meter on, insert test strip as shown. Step 2: Press the Back or Forward button one time to enter the control solution mode. The control solution bottle will appear in upper right portion of the screen. Step 3: a. Gently mix Control Solution by inverting three times b. Discard first drop c. Apply second drop to cap

3

4

Step 4:  Bring meter and strip to drop. Test strip will draw up the solution. Step 5:  Compare result to the range printed on test strip bottle.

5

Date Opened (Fecha de Apertura)

(Use within 90 days of first opening) (Usar dentro de los 90 dias a partir de la fecha en que fue abierto)

446 11182A 2010-01

CODE

Control Solution Range (mg/dL) (Rango de la solución de control)

Level 1: Level 2:

80-120 200-300

Step 6: Push strip release button forward and discard strip.

Minneapolis, MN 55439 • 800.818.8877 • www.arkrayusa.com

6

mg/dL

Indicates control solution test is needed if QC Reminder is turned on.